THE APPLICATION OF SHIFT METHOD USING MOVING LASER ON CANCER CASES IN HEAD AREA USING 3DCRT AND VMAT TECHNOLOGY

Authors

  • Samsun Samsun Department of Radiodiagnostic and Radiotherapy Engineering, Health Polytechnic of Jakarta II, Jl. Hang Jebat III Block F3 Jakarta Selatan 12720
  • Nursama Heru Apriantoro Department of Radiodiagnostic and Radiotherapy Engineering, Health Polytechnic of Jakarta II, Jl. Hang Jebat III Block F3 Jakarta Selatan 12720
  • Rudi Kurniawan Department of Radiodiagnostic and Radiotherapy Engineering, Health Polytechnic of Jakarta II, Jl. Hang Jebat III Block F3 Jakarta Selatan 12720
  • Agnes Dwika Septiti Department of Radiodiagnostic and Radiotherapy Engineering, Health Polytechnic of Jakarta II, Jl. Hang Jebat III Block F3 Jakarta Selatan 12720

DOI:

https://doi.org/10.21009/SPEKTRA.043.05

Keywords:

moving laser, verification, on-board imager

Abstract

The accuracy of patient position is essential to achieve the purpose of radiotherapy. Therefore, before radiotherapy, the isocenter of the patient must be determined to find out the target location. Verification serves to determine the isocenter that shifts from the reference point on the CT simulator. This research was conducted using field observation method, comparing the results of the isocenter after verification with the isocenter on the moving laser. Eight (8) samples from primary data were taken. Shift data was obtained from the average value and standard deviation. The instrument used was a linear accelerator of clinic iX type and On-Board Imager. The results obtained were the average lateral coordinate point of 0.055 cm; at the longitudinal isocenter point of 0.10 cm; and at the vertical isocenter point of 0.053 cm. From this research, it was concluded that the average value o the isocenter shift was less than 0.3 cm. This method can be used because the shift does not exceed the tolerance limit.

Downloads

Published

2019-12-31

How to Cite

Samsun, S., Apriantoro, N. H., Kurniawan, R., & Septiti, A. D. (2019). THE APPLICATION OF SHIFT METHOD USING MOVING LASER ON CANCER CASES IN HEAD AREA USING 3DCRT AND VMAT TECHNOLOGY. Spektra: Jurnal Fisika Dan Aplikasinya, 4(3), 143–148. https://doi.org/10.21009/SPEKTRA.043.05