IMPLEMENTATION OF TRS-398 PROTOCOL IN ROUTINE CALIBRATION OF LINAC BY DETERMINATION OF SLAB PHANTOM ON WATER PHANTOM CORRECTION FACTOR
DOI:
https://doi.org/10.21009/SPEKTRA.072.03Keywords:
dosimetry, slab phantom, water phantom, IAEA TRS-398, LINACAbstract
The water phantom is used for LINAC calibration to measure absorbed dose radiation. Practically, it requires a long preparation time and is considered less efficient. To increase efficiency, the medical physics team in a hospital uses slab phantom as the calibration tool. Consequently, the correction factor is crucial to define the equivalence of the absorbed doses resulted from slab phantom. The absorbed dose measurement was performed according to the IAEA TRS-398 dosimetry protocol with a cylindrical ionization chamber detector for 6 MV photon beam and electron beams from Elekta Synergy Platform 154029 LINAC with 6 MeV, 8 MeV, 10 MeV, and 12 MeV energy variations. The field size for slab and water phantom is 30 cm x 30 cm x 30 cm. Based on the TRS-398 protocol, the correction factor of the slab phantom calculated based on absolute dosimetry for 6 MV photons beam, the electron beam of 6 MeV, 8 MeV, 10 MeV, and 12 MeV are 1.0018; 0.9995; 0.9979; 1.0041 and 1.0068, respectively. As a result, the absorbed dose radiation measured by the calibrated slab phantom using the resulted correction factor has an equivalent amount to the water phantom.
References
[2] N. Ade, D. Van Eeden and F. C. P. du Plessis, “Characterization of Nylon-12 as a water-equivalent solid phantom material for dosimetric measurements in therapeutic photon and electron beams,” Applied Radiation and Isotopes, vol. 155, p. 108919, 2020.
[3] B. Tiwari et al., “Tissue-equivalent dosimeters based on copper doped lithium tetraborate single crystals for radiotherapy,” Radiation Measurements, vol. 151, p. 106704, 2022.
[4] D. A. Jaffray and M. K. Gospodarowicz, “Radiation Therapy for Cancer,” In: Gelband H, Jha P, Sankaranarayanan R, Horton S, editors, Washington (DC), 2015.
[5] A. Lima-Flores et al., “Analysis and characterization of neutron scattering of a Linear Accelerator (LINAC) on medical applications,” Journal of Nuclear Physics, Material Sciences, Radiation and Applications, vol. 5, no 1, pp. 65-78, 2017.
[6] I. Rosenberg, “Radiation Oncology Physics: A Handbook for Teachers and Students,” British Journal of Cancer, vol. 98, no. 5, p. 1020, 2008.
[7] M. Bencheikh, A. Maghnouj and J. Tajmouati, “Dosimetry quality control based on percent depth dose rate variation for checking beam quality in radiotherapy,” Reports of Practical Oncology and Radiotherapy, vol. 25, no. 4, pp. 484-488, 2020.
[8] S. Yazdani, F. S. Takabi and A. Nickfarjam, “The commissioning and validation of eclipseTM treatment planning system on a varian vitalbeamTM medical linear accelerator for photon and electron beams,” Frontiers in Biomedical Technologies, vol. 8, no. 2, pp. 102-114, 2021.
[9] H. R. Baghani, M. Robatjazi and S. Andreoli, “Comparing the dosimeter-specific corrections for influence quantities of some parallel-plate ionization chambers in conventional electron beam dosimetry,” Applied Radiation and Isotopes, vol. 179, p. 110031, 2022.
[10] J. Renaud et al., “Absorbed dose calorimetry,” Physics in medicine and biology, vol. 65, no. 5, p. 05TR02, 2020.
[11] H. R. Baghani, S. Andreoli and M. Robatjazi, “Performance characteristics of some cylindrical ion chamber dosimeters in Megavoltage (MV) photon beam according to TRS-398 dosimetry protocol,” Radiation Physics and Chemistry, vol. 180, pp. 109299, 2021.
[12] P. Andreo et al., “IAEA Technical Report Series No.398,” Absorbed Dose Determination in External Beam Radiotherapy, International Atomic Energy Agency, Vienna, vol. 398, 2006.
[13] H. R. Baghani and M. Robatjazi, “Scaling factors measurement for intraoperative electron beam calibration inside PMMA plastic phantom,” Measurement: Journal of the International Measurement Confederation, vol. 165, p. 108096, 2020.
[14] G. Martin-Martin et al., “Assessment of ion recombination correction and polarity effects for specific ionization chambers in flattening-filter-free photon beams,” Physica Medica, vol. 67, pp. 176-84, 2019.
[15] S. A. Pawiro et al., “Modified electron beam output calibration based on IAEA Technical Report Series 398,” Journal of applied clinical medical physics, vol. 23, no. 4, p. e13573, 2022.
[16] M. Pimpinella, L. Silvi and M. Pinto, “Calculation of k Q factors for Farmer-type ionization chambers following the recent recommendations on new key dosimetry data,” Physica Medica, vol. 57, pp. 221-230, 2019.
[17] J. Madamesila et al., “Characterizing 3D printing in the fabrication of variable density phantoms for quality assurance of radiotherapy,” Physica Medica, vol. 32, no. 1, pp. 242-247, 2016.
[18] A. Pehlivanlı and M. H. Bölükdemir, “Investigation of the effects of biomaterials on proton Bragg peak and secondary neutron production by the Monte Carlo method in the slab head phantom,” Applied Radiation and Isotopes, vol. 180, pp. 1-7, 2022.
[19] M. Bencheikh, A. Maghnouj and J. Tajmouati, “Mathematical parameterization of dosimetry quality index checking of the photon beam based on IAEA TRS-398 protocol,” Journal of King Saud University - Science, vol. 31, no. 4, pp. 1543-1546, 2019.
[20] A. Fauzi, Darmawati and F. Aliyah, “Penentuan Faktor Koreksi Slab Phantom Terhadap Water Phantom Pada Dosimetri Absolut Berkas Foton Dan Elektron Pesawat Linac Berdasarkan Iaea Trs-398,” Universitas Gadjah Mada, 2018.
[21] D. K. Bewley, “Central axis depth dose data for use in radiotherapy, A survey of depth doses and related data measured in water or equivalent media,” British journal of radiology Supplement, vol. 17, pp. 1-147, 1983.
[22] E. B. Podgorsak, “Radiation Oncology Physics: A Handbook for Teachers and Students,” IAEA. Vienna: IAEA, pp. 249-292, 2005.
[23] S. Vynckier, Heijmen et al., “Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams,” Netherland, 2012.
[24] N. Ipe, “Shielding Design and Radiation Safety of Charged Particle Therapy Facilities, PTCOG Report 1,” Particle Therapy Cooperative Group (PTCOG), 2010.
[25] E. Vano et al., “Dosimetric quantities and effective dose in medical imaging: a summary for medical doctors,” Insights into Imaging, vol. 12, no. 1, p. 99, 2021.
[26] S. Unscear, “effects of Ionizing Radiation,” United Nations, New York, pp. 453-487, 2000.
[27] K. Debertin and R. G. Helmer, “Gamma- and X-ray spectrometry with semiconductor detectors,” Netherlands: North-Holland, 1988.
[28] A. Miah et al., “Natural Radioactivity and Associated Dose Rates in Soil Samples of Malnichera Tea Garden in Sylhet District of Bangladesh,” Journal of Nuclear and Particle Physics, vol. 2, no. 6, pp. 147-152, 2013.
Downloads
Published
How to Cite
Issue
Section
License
SPEKTRA: Jurnal Fisika dan Aplikasinya allow the author(s) to hold the copyright without restrictions and allow the author(s) to retain publishing rights without restrictions. SPEKTRA: Jurnal Fisika dan Aplikasinya CC-BY or an equivalent license as the optimal license for the publication, distribution, use, and reuse of scholarly work. In developing strategy and setting priorities, SPEKTRA: Jurnal Fisika dan Aplikasinya recognize that free access is better than priced access, libre access is better than free access, and libre under CC-BY or the equivalent is better than libre under more restrictive open licenses. We should achieve what we can when we can. We should not delay achieving free in order to achieve libre, and we should not stop with free when we can achieve libre.
SPEKTRA: Jurnal Fisika dan Aplikasinya is licensed under a Creative Commons Attribution 4.0 International License.
You are free to:
Share - copy and redistribute the material in any medium or format
Adapt - remix, transform, and build upon the material for any purpose, even commercially.
The licensor cannot revoke these freedoms as long as you follow the license terms.